Home
91-2599-05 Combination pH/mV and EC/TDS/NaCl Meter
Combination pH/mV and EC/TDS/NaCl Meter for Laboratories, Hanna HI 255
HI 255 is a high performance precision instrument that measures a multitude of parameters, each over a wide range. With this single laboratory bench meter you can measure pH, ORP, conductivity, TDS, NaCl percentage and temperature.
Utilizing an external temperature probe, pH readings are automatically temperature compensated. pH calibrations are customizable - up to 3 reference points, chosen from the 5 available memorized buffers - to ensure a higher level of precision. EC measurements can be compensated relative to a selected reference temperature of 20 or 25°C. The EC calibration mode allows you to chose from among 6 memorized buffers and performed at single-point calibration. The most suitable EC and TDS range for your application is automatically selected.
With a built-in logging function, measurements are stored in non volatile memory, and can be downloaded to a PC through the RS232 serial port. In addition to these features, HI 255 is contains GLP features that store the calibration and measurement data of any sample, to be displayed upon request.
Specifications
|
Range
|
pH/ORP
|
-2.00 to 16.00 pH; -2.000 to 16.000 pH / ±699.9 mV; ±2000 mV
|
|
Range
|
EC
|
0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 mS/cm
3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual EC*
|
|
Range
|
TDS
|
0.00 to 14.99 ppm; 15.0 to 149.9 ppm; 150 to 1499 ppm;
1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L actual TDS*
|
|
Range
|
NaCl
|
0.0 to 400.0% NaCl
|
|
Range
|
Temperature
|
-10 to 120.0°C (pH); 0.0 to 60.0°C (EC)
|
|
Resolution
|
pH/ORP
|
0.01 pH; 0.001 pH / 0.1 mV (±699.9 mV); 1 mV (±2000 mV)
|
|
Resolution
|
EC
|
0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
|
|
Resolution
|
TDS
|
0.01 ppm; 0.1 ppm; 1 ppm; 0.01 g/L; 0.1 g/L
|
|
Resolution
|
NaCl
|
0.1% NaCl
|
|
Resolution
|
Temperature
|
0.1°C
|
|
Accuracy
|
pH/ORP
|
±0.01 pH; ±0.002 pH / ±0.2 mV (±699.9 mV); ±1 mV (±2000 mV)
|
|
Accuracy
|
EC
|
±1% of reading ±(0.05 µS/cm or 1 digit)
|
|
Accuracy
|
TDS
|
±1% of reading ±(0.03 ppm or 1 digit)
|
|
Accuracy
|
NaCl
|
±1% of reading
|
|
Accuracy
|
Temperature
|
±0.4°C (excluding probe error)
|
|
Relative mV Offset Range
|
|
±2000 mV
|
|
Calibration
|
pH
|
automatic, at 1, 2 or 3 points, with 5 memorized buffers
(4.01, 6.86, 7.01, 9.18, 10.01)
|
|
Calibration
|
EC
|
automatic, 1 point, with 6 memorized values
(84.0 and 1413 µS/cm; 5.00, 12.88, 80.0 and 111.8 mS/cm)
|
|
Calibration
|
NaCl
|
automatic, 1 point, with HI 7037L calibration solutions
|
|
Temperature Compensation
|
|
manual or automatic, -10.0 to 120.0°C (14 to 248°F) for pH;
0.0 to 60.0°C (32 to 140°F) for EC;
(can be diabled for measuring actual EC and TDS values)
|
|
Temperature Coefficient
|
|
0.00 to 6.00%/°C (EC and TDS only); default value: 1.90%/°C
|
|
TDS Conversion Factor
|
|
0.40 to 0.80; default value: 0.50
|
|
pH Electrode
|
|
HI 1131B (included)
|
|
EC Probe
|
|
HI 76310 (included)
|
|
Temperature Probe
|
|
HI 7662 (included)
|
|
PC Connection
|
|
RS232 opto-isolated serial port
|
|
Data Logging
|
|
200 samples
|
|
Input Impedance
|
|
1012 Ohm
|
|
Power Supply
|
|
12 Vdc adapter (included)
|
|
Environment
|
|
0 to 50°C (32 to 122°F); RH max 95%
|
|
Dimensions
|
|
240 x 182 x 74 mm (9.4 x 7.2 x 2.9")
|
|
Weight
|
|
1.1 kg (2.4 lb.)
|
| Catalog # |
Product Description |
|
Price |
|
91-2599-05
|
Combination pH/mV and EC/TDS/NaCl Meter for Laboratories, Hanna HI 255
|